Determination of rare earth elements and yttrium in nine geochemical reference samples using a novel group separation procedure involving mixed-acid elution ion-exchange chromatography
I.W. Croudace and S. Marshall
Department of Geology,
The University,
Southampton SO9 5NH, U.K.
Abstract
Rare earth data for nine well-known geochemical reference samples
were obtained using a rapid and efficient cation exchange chromatographic
separation procedure. It involves the use of a mixed-acid eluent
(HCl and HNO3) on a 15 x 1 cm column of Bio-Rad AG50-X8 (200-400
mesh) to remove matrix and many trace elements from rock solutions.
The rare earth elements are removed from the column using 60 ml
7M HNO3. The pre-concentration procedure described is routinely
used in this laboratory for the determination of the rare earth
elements by mass spectrometry-isotope dilution, and has been satisfactorily
used for their determination using inductively-coupled plasma-source
atomic emission spectrometry.
Résumé
Les éléments de terres rares ont été
dosés dans neuf échantillons géochimiques
de référence en utilisant une méthode chromatographique
rapide de séparation à l'aide d'une résine
échangeuse de cations. La méthode consiste à
utiliser un éluant mixte (HCI et HNO3) sur une colonne
de 15 x 1 cm contenant la résine Bio-Rad AG50-X8 (200-400
mesh). Les éléments de terres rares sont récupérés
de la colonne avec 60 ml de 7M HNO3. Cette technique de préconcentration
de terres rares est utilisée dans ce laboratoire en analyse
de routine par spectrométrie de masse à dilution
isotopique et a été employée de façon
satisfaisante pour leur dosage par la spectrométrie d'émission
à plasma induit (ICP).