Determination of beryllium and zirconium in 45 geochemical reference samples by inductively coupled plasma emission spectrometry
P.J. Watkins and M. Thompson
Department of Geology,
Imperial College of Science and Technology,
London SW7 2BP, England
Abstract
The beryllium and zirconium contents of 45 geochemical reference
samples have been determined by inductively coupled plasma after
fusion of the samples with lithium metaborate and dissolution
of the melt in dilute nitric acid. The method described here is
rapid and sample preparation straightforward. Good agreement is
shown with previously published results for these two elements.
A correction has to be made for an interference due to vanadium
in determining the beryllium content, and there is a slight interference
due to yttrium in the determination of zirconium. The detection
limit for beryllium is about 0.2 mg g-1 and for zirconium about
15 mg g-1 in the sample.
Résumé
La teneur en béryllium et en zirconium dans 45 échantillons
géochimiques de référence a été
dosée par spectrométrie d'émission à
plasma induit. L'échantillon a été mis en
solution par fusion avec du métaborate de lithium suivie
de la dissolution avec de l'acide nitrique dilué. La procédure
décrite est simple et facile à mettre en oeuvre.
Les résultats obtenus montrent une bonne concordance avec
des valeurs publiées pour ces deux éléments.
I1 a fallu faire une correction pour le dosage du béryllium
à cause de l'interférence spectrale due au vanadium;
le dosage du zirconium est affecté par une interférence
due à l'yttrium mais dans un moindre degré. La limite
de détection du béryllium est environ 0.2 µg
g-1 et pour le zirconium 15 µg g-1 dans l'échantillon.
Geostandards Newsletter
(1983), Vol. 7 No. 2 pp. 273-277